EQUIPMENT    Laryngoscope  Troubleshooting
Laryngoscope  
Flowchart  


     




Description  #   Text  Box   Comments  
1   Start:  Laryngoscope  Troubleshooting           Begin  diagnostic  process  for  a  work  order  for  Laryngoscope.  
2   Is  the  laryngoscope  free  of  any  external  damage/defects?   Inspect  laryngoscope  for  external  cracks,  damage,  or  dirt.  
3   Rinse  blade  under  tap  water  and  disinfect/sterilize.   Clean  and  dry  the  blade.  
4   Clean  any  build-­‐up  of  material  on  contact  strips.                                                 Examine  the  contact  strips  (blade-­‐handle  interface)  of  laryngoscope  for  dirt.    
5   Remove  any  peeling  chrome  (from  blade)  with  a  scalpel  blade  and  rub  over  the  edge  with  sandpaper.    


Examine  the  blade  for  any  peeling  chrome  and  remove  peels  with  care.  Ensure  that  there  is  no  damage  to  the  blade  during  the  removal  process.  
6   Does  the  light  bulb  power  on  with  battery?   Power  on  the  device  with  battery.  Examine  whether  light  bulb  gives  sufficient  illumination.  
7   Recharge  battery  if  applicable  or  replace  it.           Refer  BTA  skill  set  on  Batteries  to  identify  and  replace  damaged  batteries.  
8   Replace  bulb  if  necessary.                       Refer  BTA  skill  set  on  Lighting/Indicators  to  replace  non-­‐functional  light  bulbs.  
9   Inspect  and  fix  broken  wires  or  bad  connections  between  the  battery  and  bulb.                                        


Inspect  wires  and  connections  from  battery  to  bulb  using  multimeter.  Refer  BTA  skill  set  on  Connections  for  identifying  and  fixing  broken  wires  and  bad  connections.  
10   Does  the  device  use  a  fiber  optic  cable  and  still  the  light  bulb  doesn't  power  on  with  battery?    


There  are  two  types  of  laryngoscopes:  
• Conventional  laryngoscope  • Fiber  optic  laryngoscope  If  the  bulb  still  doesn’t  power  on  then  there  might  be  cracks  in  the  fiber  optic  strands.  So  the  unit  might  have  to  be  replaced.  


11   Inspect  for  cracks  in  fiber  optic  strands  and  replace  the  unit  if  necessary.               If  the  illumination  is  not  sufficient  then  inspect  for  cracks  in  fiber  optic  strands.  
12   Is  the  lens  free  from  cracks/defects?   Any  defect  in  the  lens  obstructs  vision  of  larynx  and  vocal  cords.  
13   Replace  lens  if  necessary.             Identify  and  replace  with  a  suitable  lens.  




14   Is  the  blade  and  handle  connected  properly?   Improper  connection  between  blade  and  handle  may  result  in  faulty  operation  of  the  device.  
15   Refer  device  manual  to  assemble  the  parts.   Follow  instructions  in  the  device  manual  to  assemble  the  parts.  
16   Perform  preventive  maintenance  on  Laryngoscope.  Return  device  to  clinical  personnel.  


Laryngoscope  is  working  properly.  Perform  preventive  maintenance  before  returning  the  device  to  clinical  personnel.    
 