Microsoft Word - EQU - Infant warmers PM and flowchart.doc

Troubleshooting  Flowchart  for  Infant  Warmer  


 
#   Text  Box   Comment  or  Explanation  1   Begin:  Infant  warmer   Begin  diagnostic  process  for  infant  warmer.  2   Does  warmer  power  on?   Lights,  displays,  and  sounds  are  all  indications  that  the  device  has  powered  on.  3   Troubleshoot  power  supply  (separate  flowchart)   Most  infant  warmers  have  AC-­‐DC  power  supplies.  4   Change  battery  if  necessary.   Many  infant  incubators  have  batteries.  5   Does  unit  heat  when  switched  on?   Does  the  device  produce  heat  when  the  warmer  is  switched  on?  6   Does  power  reach  the  heating  element?   If  no  heat  is  produced,  test  if  the  proper  voltage  is  reaching  the  heating  element.  7   Repair  any  open  circuits  or  bad  connections.   If  power  does  not  reach  the  heating  element,  there  might  be  an  open  circuit,  bad  connection,  or  broken  wire.  




8   Replace  switch,  relay,  or  triac.   A  switch,  relay,  or  triac  may  be  used  to  control  current  flow  to  the  heating  element.  Ensure  it  is  working  properly.  9   Is  heating  element  resistive?   Infant  warmers  typically  use  a  resistive  element  or  a  bulb  to  create  heat.  Check  which  one  is  used  in  this  device.  10   Replace  bulb.   If  a  bulb  is  used  the  proper  voltage  is  reaching  the  bulb,  it  must  be  replaced.  
11   Replace  heating  element.   If  a  resistive  element  is  used  and  the  proper  voltage  is  reaching  the  resistor,  the  element  must  be  replaced.  Resistive  elements  can  generally  be  replaced  with  any  resistor  or  combination  of  resistors  with  the  same  resistance  and  power  ratings  as  the  one  being  replaced.  
12   Is  the  warmer  designed  to  use  a  temperature  probe?   If  the  warmer  heats,  it's  necessary  to  check  the  temperature  control  methods  of  the  device.  Is  a  temperature  probe  (thermometer)  used,  either  on  the  skin  of  the  infant  or  in  the  air?  13   Is  temperature  >  40  degrees  C?   If  there  is  no  thermometer,  the  temperature  must  be  measured  at  patient  level  to  ensure  it  stays  at  the  desired  level  (typically  between  34  and  40  degrees  C).  14   Is  temperature  <  34  degrees  C?   If  the  termperature  is  below  40  degrees  C,  confirm  that  is  above  the  mimimum  temperature  (34  degrees  C).  
15   Increase  temperature.   Increase  the  temperature  if  it  is  below  34  degrees  C.  This  might  be  accomplished  by  using  control  knobs  to  increase  heat  output,  moving  the  heating  element  closer  to  the  patient,  and/or  reducing  ventilation.  
16   Decrease  temperature.   Decrease  the  temperatere  if  it  is  above  40  degrees  C.  This  might  be  accomplished  by  using  the  control  knob  to  decrease  output  of  the  heating  element,  moving  the  element  farther  from  the  patient,  and/or  increasing  ventilation.  
17   Is  probe  present?   Confirm  that  the  probe  is  present.  Missing  or  damaged  temperature  probes  are  a  common  problem.  Some  warmers  may  have  a  manual  mode  to  set  a  fixed  heat  output  if  the  temperature  probe  is  missing.  18   Is  probe  thermister?   If  there  is  a  temperature  probe  present,  it  will  be  either  a  thermistor  or  a  thermocouple.  Attempt  to  verify  which  it  is.  
19   Replace  probe  or  improvise  thermocouple.  


If  the  probe  is  a  thermocouple,  it  works  by  providing  a  voltage  source  that  varies  in  response  to  the  input  temperature.  A  battery  and  voltage  divider  might  be  able  to  be  used  to  improvise  the  voltage  output  at  a  particular  level  to  convert  the  infant  warmer  to  manual  mode.  
20   Replace  probe  or  improvise  thermister  


If  the  probe  is  a  thermister,  it  works  by  providing  a  resistance  that  varies  in  response  to  the  input  temperature.  A  potentiometer  or  resistor  might  be  able  to  be  used  to  improvise  the  resistance  at  a  particular  level  to  convert  the  infant  warmer  to  manual  mode.  
21   Does  warmer  maintain  temp  within  1  degree  C?   If  the  probe  is  present,  test  the  temperature  output  at  various  temperature  levels  to  ensure  the  device  can  maintain  the  selected  temperature  within  1  degree  C.  22   Replace  or  correct  probe.   If  the  device  does  not  maintain  a  temperature  close  to  the  input,  the  probe  must  be  replaced  or  corrected.  23   Go  to  begin.   Begin  the  diagnostic  process  again  to  determine  if  corrective  measures  have  repaired  the  device.  24   Go  to  begin.   Begin  the  diagnostic  process  again  to  determine  if  corrective  measures  have  repaired  the  device.  




25   Infant  warmer  is  working  properly.   The  infant  warmer  maitains  the  appropriate  temperature.  The  repair  was  successful.  26   Is  temperature  between  34  and  40  degrees?   Does  the  warmer  maintain  an  appropriate  temperature  (between  34  and  40  degrees  C)  when  operating  in  manual  mode  (without  a  temperature  probe)?    