Ultrasound  Repair  and  Troubleshooting    
 


 


Flowchart  


 


   




Description  


1. Begin Ultrasound Flow Chart Begin Ultrasound Flow Chart


2. Does it turn on? Does it turn on?


3. Troubleshoot Power Supply See power supply flowchart


4. Does the screen turn on? Does the screen turn on?


5. Check circuit components Troubleshoot the circuit (see troubleshooting guide)


6. Does the probe work? Does the machine show an image?


7. Check gel. Is there enough of the proper gel.


8. Check probe. Ensure that the probe is properly
connected and undamaged.


9. Repair or replace probe. Solder the broken connections or
replace the probe.


10. Do the dials work? Do the dials properly control the
machine?


11. Replace dials or use anticorrosive
contact spray on dials.


Replace or use anticorrosive contact
spray on dials.


12. Is there a flickery or partial image
Does the image on the screen have
missing sections? Does the screen
flicker?


13. Check probe.


Gently pull on the cable at different
points to see when the image flickers.
This is where the cable is broken.


For composite probes wave a pencil
across the transducer. If not seen, this
is where the transducer is not




connected.


14. Repair or replace probe. Solder the broken connections or replace the probe.


15. Go to begin. Go to begin.


16. Is there a noisy signal? Is there distortion in the produced
image?


17. Replace motor or bearings Replace the motor or bearings if they
are broken or damaged.


18. Is there major image
deterioration? Is the image severely deteriorated?


19. Are there air bubbles in the probe? Does the oil in the probe have air bubbles in it?


20. Replace oil. Replace oil.


21. Is there a sudden worsening
quality of the image?


Does the image have sudden changes
in quality?


22. Check the preamplifier. Check the preamplifier.


23. Check the mains voltage range. Check the outlet voltage range.


24. Check that the probe is connected
securely.


Check that the probe is connected
securely.


25. Is there a random change in
greyscale quality?


Is there a random change in greyscale
quality?


26. Check the monitor and the
preprocessing unit.


Check the monitor and the
preprocessing unit.


27. Can you change programs or make
measurements on the machine?


Can you change programs or make
measurements on the machine?




28. Check the digital processor. Check the digital processor.


29. Is there a partial image despite an
intact probe?


Is there a partial image despite an
intact probe?


30. Check the digital memory and the
microprocessor.


Check the digital memory and the
microprocessor.


31. Is the image wavy? Is the image wavy?


32. Is the machine calibrated
correctly?


Is the machine calibrated correctly?


33. Go to begin. Got to begin.


34. Calibrate the machine. Calibrate the machine.


35. End End  