Microsoft Word - Laryngoscope Preventative Maintenance.docx

EQUIPMENT    Microscope  Repair  and  Troubleshooting  
Microscope  Flowchart  


 




Description  #   Text  Box   Comments  
1   Begin:  Microscope   Begin  diagnostic  process  for  a  work  order  for  Microscope.  Maintenance  is  generally  requested  on  a  microscope  when  a  specimen  cannot  be  viewed  clearly  or  at  all.  
2   Does  the  microscope  power  on  when  plugged  in?   When  plugged  in,  the  microscope  should  power  on  completely.    
3   Troubleshoot  power  supply  (separate  chart).   If  no  power  reaches  the  machine,  there  may  be  problems  with  the  switch,  fuse,  power  supply  components,  or  wiring.    See  flowchart  on  Power  Supply  and  BTA  skills  on  Power  Supply.  
4   Does  the  light  source  turn  on  and  stay  on?   The  light  source  should  remain  constant  across  the  stage  when  on.  
5   Does  the  light  turn  on  and  flicker?   If  the  light  turns  on  but  does  not  remain  constant,  there  may  be  a  minor  problem  that  can  be  fixed  without  replacing  the  bulb  completely.  
6   Ensure  contact  is  clean  and  secure  any  loose  connection.   The  inside  of  the  lamp  house  or  the  connections  may  be  dirty.  The  connections  should  also  be  secured  firmly.  See  BTA  skills  for  Connections.  
7   Is  resistance  at  each  end  of  the  filament  0  ohms?   This  checks  if  the  filament  in  the  bulb  is  functioning.  
8   Check  internal  circuitry,  dimmer  knob,  or  diaphragm,  if  applicable.   Ensure  that  the  circuitry  that  connects  to  the  light  source  is  intact.  In  addition,  ensure  the  dimmer  knob  is  turned  on  and  that  the  diaphragm  is  open.  
9   Replace  bulb  with  same  type  of  bulb.  


If  the  light  source  still  does  not  turn  on,  replace  the  bulb  with  another  of  the  same  type.  If  the  same  type  of  bulb  is  not  available,  a  new  source  can  be  wired  in.  See  BTA  skills  on  Replacement  of  Light  Bulbs  and  Light  Fixtures.  
10   Go  to  begin.   Restart  the  diagnostic  process.  
11   Focus  the  microscope  at  the  lowest  objective.   Begin  this  portion  of  the  diagnostic  process  at  the  lowest  objective.  
12   Can  the  specimen  be  viewed  clearly?   If  the  specimen  is  out  of  focus  or  cannot  be  viewed  at  all,  the  objectives  may  need  attention.  
13   Clean  dirty  objective  using  a  soft  brush,  a  lint-­‐free  cloth,  and  lens  cleaner.  


Objectives  can  be  unscrewed  or  removed  for  cleaning.  Unscrew  eyepiece  to  use  as  a  magnifying  glass  if  one  is  not  available  to  inspect  objective  for  scratches,  nicks,  cracks,  deterioration  of  seal  around  lens  or  oil  seepage  into  lens.  Blow  off  any  dust  with  canned  air  before  cleaning.  Start  cleaning  objective  lens  from  




center  and  spiral  out  with  cotton  swab  or  cloth.  If  lens  cleaner  is  not  available,  ethyl  ether,  xylene,  petrol  can  be  used.  Alcohol,  acetones  or  any  other  ketones  should  not  be  used,  as  they  may  dissolve  sealants.    
14   Do  the  mechanical  adjustment  knobs  turn  easily?   The  knobs  and  stage  should  be  able  to  move  freely  and  also  maintain  a  steady  position.  The  screws  holding  each  in  place  may  need  some  adjustment.  
15   Ensure  the  correct  amount  of  lubricant  has  been  used.  


Clean  off  excess  lubricant  (especially  if  it  has  dried  and  is  clumping)  using  a  soft  cloth  dampened  with  alcohol.  Do  not  use  solvents  that  leave  residue  or  lint  on  the  surface.  Dust,  clean  (with  solvent  listed  above),  polish  (with  metal  polish,  if  available)  and  lubricate  adjustments  if  they  are  stuck  or  difficult  to  turn.  
16   Ensure  the  microscope  is  covered  to  avoid  dust  settling  on  it.     When  covering  the  microscope,  put  a  small  amount  of  uncooked  rice  to  prevent  fungal  growth.  Replace  uncooked  rice  weekly.  
17   Microscope  is  working  properly.   Return  the  machine  to  the  appropriate  clinical  personnel.  
 
 




Copyright 2016, Engineering World Health