Troubleshooting  Flowchart  for  Bililight    


 #   Text  Box   Explanation  or  Comment  1   Begin:  Bililight   Start  the  diagnostic  process  for  a  work  order  on  a  bililight.  2   Does  device  power  on?   Displays,  lights,  and  sounds  are  all  indicators  that  the  device  has  turned  on.  3   Troubleshoot  power  supply  (separate  flowchart).   Most  bililights  have  an  AC-­‐DC  power  supply.  4   Change  battery  if  necessary.   Some  bililights  may  have  a  batterty.  5   Do  lights  come  on?   Do  the  blue  lights  illuminate  when  they  are  switched  on?  6   Does  power  reach  bulb?   Does  the  appropriate  voltage  reach  the  input  for  the  bulb?  7   Revise  intensity  control  knob.   Make  sure  any  intensity  control  knob  is  turned  up  to  allow  the  lights  to  illuminate.  8   Check  for  bad  switches  and  knobs.   Review  switches  and  potentiometers  for  faults.  9   Check  for  bad  connections  and  open  circuits   Open  circuits,  bad  connections,  and  bad  wires  can  prevent  the  lights  from  illuminating.  




10   Ensure  device  is  not  overheating.  Repair/maintain  fans  and  filters  if  necessary.  
Some  bililights  have  a  safety  mechanism  that  turns  the  lights  off  when  the  device  overheats.  If  this  is  the  case,  try  allowing  the  device  to  cool  before  attempting  to  turn  the  lights  on  again.  If  this  was  the  problem,  consider  ways  to  improve  ventilation.  


11   Replace  bulbs.   If  the  appropriate  voltage  reaches  the  bulbs  and  they  do  not  illuminate,  the  bulbs  must  be  replaced.  
12   Is  light  intensity  >  4μW/cm2/nm   Measure  the  light  output  at  patient-­‐level.  4μW/cm2/nm  is  the  minimum  output  for  a  bililight,  though  many  have  outputs  higher  than  this  (e.g.  15μW/cm2/nm)  
13   Reposition  light  source.   If  the  intensity  is  too  low,  the  light  source  may  be  position  too  far  from  the  patient.  14   Revise  intensity  control  knob.   The  light  intensity  might  be  increased  by  manipulating  the  intensity  control  knob.  15   Clean  interior  of  machine.   If  the  interior  of  the  machine  or  the  bulbs  are  dirty,  it  might  obstruct  the  light  source  or  cause  overheating  of  the  device.  16   Go  to  begin.   Start  the  diagnostic  process  again  to  see  if  the  corrective  measures  have  solved  the  problem.  17   Bililight  is  working  properly.   The  light  intensity  is  above  the  minimum,  and  the  device  can  be  used  with  patients.    