Defibrillator  Troubleshooting  and  Repair  


 
Description  




  Text  Box   Comments  1   Start:  Defibrillator  Troubleshooting.         Begin  diagnostic  process  for  a  work  order  on  Defibrillator  
2   Is  the  defibrillator  free  of  any  external  damage/defects?   Inspect  defibrillator  for  external  cracks,  broken  switch,  knobs  and  indicators.  
3   Identify  and  replace  damaged  switches,  indicators.   See  BTA  skill  set  on  Switches  and  Lighting/Indicators  to  identify  and  replace  damaged  switches  and  indicators.  
4   Clean  casing,  pads  using  Alcohol.   Examine  casing,  pads  and  cables  of  defibrillator  for  gel  and  dirt.  Refer  BTA  skill  set  on  Cleaning  to  clean  the  defibrillator.  If  necessary,  address  damage  to  casing  with  BTA  skills  on  Casing.  
5   Does  the  defibrillator  power  on  from  ac  power  line?   Power  the  device  from  ac  line  and  turn  it  on.  
6   Inspect  AC  adapter  cable  for  cuts,  broken  wires  and  replace  if  necessary.   See  BTA  skill  set  on  Connections  and  Connectors  for  identifying  and  replacing  damaged  cables.  
7   Inspect  and  fix  broken  wires  or  bad  connections  inside  the  defibrillator.  


Inspect  wires  and  connections  from  power  supply  circuit  board  to  other  boards  using  multimeter.  See  BTA  skill  set  on  Connections  for  identifying  and  fixing  broken  wires  and  bad  connections.  
8   Identify  and  replace  blown  fuse.   See  BTA  skill  set  on  Fuse  to  identify  and  replace  blown  fuse.  
9   Troubleshoot  power  supply.   Most  defibrillators  can  power  on  from  battery  and  ac  power  mains.  See  flowchart  on  Power  Supply,  and  BTA  skills  on  Power  Supply.  
10   Does  the  defibrillator  power  on  with  battery?  


Disconnect  defibrillator  from  ac  power  line.  Turn  the  device  on.  
• If  battery/status  indicator  is  red  then  battery  needs  to  be  charged  or  replaced  (non-­‐rechargeable).  • If  defibrillator  fails  to  power  on  then  battery  is  fully  depleted  or  damaged.  11   Recharge  battery  if  applicable  or  replace  it.   See  BTA  skill  set  on  Batteries  to  replace  and  identify  damaged  batteries.  


12   Troubleshoot  charging  circuit  if  battery  doesn't  or  charges  very  slowly.   See  BTA  skill  set  on  Transformer  and  Regulators  to  troubleshoot  charging  circuit.  
13   Run  the  defibrillator  self-­‐test.   Power  the  device  from  ac  line  and  turn  it  on.  The  device  will  run  an  automatic  self-­‐test.  
14   Does  display  show  error  or  status  indicator  is  red?   The  result  of  the  self-­‐test  will  be  displayed  (on  the  screen)  or  status  indicator  will  change  red/green.  




 

15   Are  the  defibrillator  paddles  and  cables  damaged,  wet?   The  paddles  should  be  clean  and  dry.  Inspect  the  pad  cables  and  connectors  for  cuts  and  broken  wires.  
16   Does  the  defibrillator  use  paddle  electrodes?   Paddle  electrodes  consist  of  a  metal  paddle  with  an  insulated  handle.  
17   Clean  paddles  using  alcohol.  Identify  and  replace  damaged  paddles  and  cables.  


Paddle  electrodes  are  reusable  and  should  be  cleaned  after  every  use.  See  BTA  skill  set  on  Connections  and  Connectors  for  identifying  and  replacing  damaged  cables.  
18   Replace  the  Self-­‐adhesive  electrode.   Self-­‐adhesive  electrodes  should  be  replaced  after  every  use.  
19   Does  the  hospital  have  a  defibrillator  tester?   Defibrillator  testing  can  be  done  on  a  commercial  tester  or  a  large  piece  of  meat.  
20   Test  defibrillator  using  the  defibrillator  tester.  


Connect  pads  to  defibrillator  analyzer.  Select  energy  and  press  charge  button.  Once  charged  push  discharge  button.  Record  delivered  energy  from  display  of  defibrillator  analyzer.  Repeat  the  procedure  for  different  energy  levels.  
21   Is  delivered  energy  >=  limit  set  by  manufacturer  (refer  manual)  for  different  energy  levels?  


Improper  functioning  of  internal  circuitry  if  the  defibrillator  delivers  less  or  no  energy  than  the  limit  set  by  manufacturer.    
22   Test  the  defibrillator  using  a  large  piece  of  meat.  


Set  energy  to  maximum  and  press  charge  button.  Once  charged  place  pads  on  a  large  piece  of  meat.  Press  discharge  button.  Repeat  the  procedure  10  times.  The  piece  of  meat  should  be  large  enough  so  that  the  defibrillator  paddles  can  be  placed  greater  than  6  inches  apart.  
23   Are  there  no  burn  marks  on  the  meat  (after  10  shocks)?   Failure  of  internal  circuitry  if  no  burn  marks  are  found  on  the  piece  of  meat.  
24   Refer  equipment  to  device  manufacturer.          Refer  equipment  to  device  manufacturer  for  possible  repair  and  replacement  of  internal  circuitry  components.  
25   Perform  preventive  maintenance  on  defibrillator.  Return  defibrillator  to  clinical  personnel.                          


Defibrillator  is  working  properly.  Perform  preventive  maintenance  before  returning  the  device  to  clinical  personnel.    