EQUIPMENT    Oxygen  Concentrator  Preventative  Maintenance  
Preventive  Maintenance    Always  run  machine  for  a  few  hours  after  maintenance  before  patient  use  


1. Clean  humidifier  and  tubes  a. Wash  in  warm  soapy  water,  rinse  thoroughly,  and  replace  2. Change  distilled  water  3. Remove  and  clean  the  foam  air  intake  filter    a. Wash  with  soap  and  water  b. Ensure  filter  is  dry  before  replacing  but  do  not  use  heat  to  dry  4. Check  Alarm  system  battery  a. An  audible  alarm  should  sound  when  the  machine  is  turned  on  b. Replace  9V  battery  when  necessary  5. Check  if  Zeolite  Canisters  have  expired  (25,000  hours).  The  granules  become  grey  when  they  are  no  longer  effective.    6. Felt  pre-­‐filter  should  be  changed  once  a  month  (if  available)  7. Replace  patient  bacterial  filter  annually  (if  available)  
 
 